File name ReleaseNotes_SW ReleaseNotes_SW_Desktop_A050220121101.txt== Agilent B1500A/B1505A Software A.05.02.2012.1026 Release Notes ==
== CORRESPONDING FIRMWARE REVISION ==
This software A.05.02.2012.1026 works with the following firmware.
- B1500A/B1505A: A.05.02 or later.
- 415xB/C: HOSTC 03.08, SMUC 04.08 or later.
== FIXED DEFECTS ==
Following defects are fixed in this release A.05.02.2012.1026
- [8839] [Oscilloscope View] Time domain may not be set to 5us/div by enter key.
- [8850] [C-V Sweep] For imported C-V Sweep setup, CONST SMU channel source setting may be reset to 0 depending on the order in channel definition.
- [8852] [N1267A Support] 0V is output for Gate instead of GateOff setting for FET Current Collapse Signal Monitor Application Test.
- [8855] [Classic Test] Channel Setting in Advanced Setup dialog may miss and duplicate SMU channels depending on GNDU order in Channel Definition.
- [8856] [Database] database may corrupt by saving App Test Definition during measurement.
- [8858] [EasyEXPERT for 415x] Error may occur with execution Application Test imported from Test Setup file.
- [8862] SMU/PG Selector setting and ASU I/O Path setting are not restored at the opening of a worksapce.
- [8864] Recovery from non-numeric input for numeric field needs keyboard.
- [8866] [Auto/Folder Export] Access Control List of index.txt is not derived from the target folder.
- [8868] [N1268A Support] Diagnosis may end up with timeout error in case of wrong V Control SMU connection.
- [8870] [Oscilloscope View] Sampling setup is disabled in case that only arithmetic operation data is selected.
- [8873] [Tracer Test] Axis setting dialog may not appear appropriate position in case of Desktop version.
- [8904] [Tracer Test] User Function converted from arithmetic operation with right-hand side operand of signed literal constant is invalid.
== REVISION HISTORY ==
New Features:
- N1267A support for B1505A
- N1265A diagnosis
- N1268A diagnosis
Fixed Defects:
- [8756] [Tracer Test] Specified minimum value of scale may not be restored from my favorite setup.
- [8779] [N1265A Support] Thermo-trigger does not refer latest temperature.
- [8783] [Database Restoration] Access Permission of Database changes.
- [8786] [Oscilloscope View] Error dialog persists for time scale input of wrong format.
- [8789] [Multi Channel I/V Sweep] Incorrect Japanese error message for error 105940.
- [8790] [B1511A Support] MPSMU is to be supported in Application Tests for B1505A.
- [8796] [N1265A Thermometer Support] Thermo-trigger setting GUI on Repeat Measurement Setup is to be disabled in case of other than B1505A.
- [8802] [N1265A Support] Performing self-test resets default output.
- [8815] [Oscilloscope View] Time domain scale setting dialog persists after switching to Test Modes other than Tracer Test Mode.
- [8824] [N1265A Support] Wrong error handling |